WINTERHALTER SOFTWARE DOWNLOAD
Winterhalter constantly strives to improve its products and periodically provides updates for the software. To keep your machine running smoothly and delivering the best attended results, update the software on a regular basis.
Important Note
From 6 January 2025, the following software packages can only be installed if the date on the machine is set to a date before 6 January 2025.
• UC V02: UC000093
• PT V02: PT000055
• CTR: CTR000040
The reason for this is that the signature of the software packages will no longer be valid from 6 January 2025. However, as the signature is only checked during installation, the current date can be set again after installation.
In order to install higher software packages, the software package listed above must always be installed first if a lower software version is installed on the machine. The software packages listed above are therefore permanently available.
Current Update
RootFS version: 4.3
HMI (app) version: 3.1.3
Frequency inverter version: V73
Control unit version: 1.4.1
Control unit mini version: 1.5.0
Conductivity measurement unit version: 1.0.0
Quick-start button version: 1.1.0
The update contains the following changes:
- Door safety circuits integrated into the actuator programme (inputs DS.1 and DS.2).
- Error messages ERR 191 - ERR 193 for errors in the door safety circuits were added.
- Error message ERR 57 was added. It indicates a blocked funnel on the SPI solid dosing device.
- Sensitivity of error message ERR 904 for machines with heat pump has been adjusted.
- The sensitivity of error messages ERR 270 - ERR 290 for machines with a drying or air knife drying zone has been adjusted.
- Evaluation of the measured temperatures in the drying or air knife drying zone has been adjusted.
- Error message ERR 69 has been added. It indicates missing flaps at the unloading zone in the MTF.
- Overload detection by the control unit has been improved.
- Error messages ERR 70 - ERR 74 have been replaced by error messages ERR 40 - ERR 44. They indicate that the tank level is too low.
- FlexSpeedEnd feature has been added. Transport speed decreases if the transport limit switch is triggered too often.
Transport speed increases if the transport limit switch is triggered less frequently. - For machines with a heat pump or EnergyPlus heat exchanger, the maximum rinse water volume is delivered for 5 seconds (P262) when the rinse pump restarts. This counteracts a drop in the amount of rinse water when operating the machine for a longer period of time.
- Overlapping texts have been removed.
RootFS version: 4.3
Frequency inverter (rack transport) version: 48
Frequency inverter (rinse booster pump) version: 48
HMI (app) version: 3.0.4
Control unit version: 1.0.10
Power unit version: 2.0.0
The update contains the following changes:
- Slow reaction of the display has been fixed.
- Temperatures in the info menu are displayed in the correct order for machines with ‘right-left’ working direction.
- Error 204 and Error 210 are only displayed if the error is present.
Please note:
- The software update can only be installed if at least software version CTR000040 is installed on the machine.
- If this is not the case, set the date on the machine to 05/01/2025 and then install software update CTR000040.
- Then set the current date and install the latest software update.
RootFS version: 4.3
Frequency inverter version: V48
HMI (app) version: 3.0.3
Control unit version: 1.2.5
The update contains the following changes:
- Slow reaction of the display has been fixed.
- Display of Error 403 is reactivated.
Please note:
- The software update can only be installed if at least software version UC000093 is installed on the machine.
- If this is not the case, set the date on the machine to 05/01/2025 and then install software update UC000093.
- Then set the current date and install the latest software update.
RootFS version: 4.3
Frequency inverter version STM32: 20
Frequency inverter version ATSAM: 22
HMI (app) version: 3.0.3
Control unit version: 1.3.2
Automatic hood software version: 1.2.0
Power unit version: 4.0.0
The update contains the following changes:
- Slow reaction of the display has been fixed.
Please note:
- The software update can only be installed if at least software version PT000055 is installed on the machine.
- If this is not the case, set the date on the machine to 05/01/2025 and then install software update PT000055.
- Then set the current date and install the latest software update.
RootFS version: 4.3
Frequency inverter (rack transport) version: 48
Frequency inverter (rinse booster pump) version: 48
HMI (app) version: 3.0.2
Control unit version: 1.0.9
Power unit version: 2.0.0
The update contains the following changes:
- Optimization and bug fixes
RootFS version: 4.3
Frequency inverter version STM32: 20
Frequency inverter version ATSAM: 22
HMI (app) version: 3.0.2
Control unit version: 1.3.2
Automatic hood software version: 1.2.0
Power unit version: 4.0.0
The update contains the following changes:
- Optimization and bug fixes
RootFS version: 4.3
Frequency inverter version: V48
HMI (app) version: 3.0.2
Control unit version: 1.2.4
The update contains the following changes:
- Optimization and bug fixes